4.3 Measurement Current Pulse Output Function
62
Measurement current application timing when measurement current pulse mode is on
Measurement
Measurement
Measurement
Measurement
current
current
Measurement current application when using normal internal trigger measurement
(The measurement current is applied continuously when measurement current pulse mode is off.)
Measurement processing triggered by external TRIG signal or "READ?" command
Applied only during measurement (including trigger delay time).
Measurement and current application when using external trigger measurement
or when continuous measurement is off (:INIT:CONT OFF)
Measurement current is applied continuously.
When using the instrument with the measurement current pulse output func-
tion enabled, always enable the trigger delay function and set the delay time
to 10 ms or more.
See "Trigger Delay Settings" (p.60)
When the measurement current pulse output function is on, the below is
added to the sampling time.
When using the V mode: +1 ms
When using the mode: +4 ms