2.4 Turning the Power On and Off
22
2.4 Turning the Power On and Off
Before turning the instrument on, make sure the supply voltage matches that
indicated on the its power connector. Connection to an improper supply volt-
age may damage the instrument and present an electrical hazard.
The measurement setting state is the same as when the power was previ-
ously turned off (backup).
To preserve changes to settings, wait a short time (about five seconds) after
changing a setting before turning power off.
However, measurement settings made through the RS-232C, LAN, or GP-
IB interface and measurement settings loaded by LOAD
signals of the EXT
I/O connector are not memorized.
Before starting to measure, allow 30 minutes for warm-up.
After warm-up, be sure to perform a self-calibration.
See "4.10 Self-Calibration" (p.73).
Turn on the main power switch on ( ) the rear of the instrument.
The instrument will start up in the standby state in which it was last
turned off. (The instrument ships in the standby state.)
Turning On ( ) the Main Power Switch (Rear of Instrument)
Power ON
Turn off the main power switch on the rear of the instrument.( ).
Turning the Power Off
Power OFF
HIOKI BT3562A981-09