4.1 Comparator Function
53
4
Chapter 4 Applied Measurement
Two methods are available for setting comparator thresholds.
Comparator Threshold Method Selection
1
(The SHIFT indicator lights up.)
The Comparator setting display appears.
2
Press so that the indicated position blinks, and set the comparator threshold
method.
HIGH, LOW..... Compare against specified upper and lower thresholds (default setting
method)
REF, % ............ Compare against upper and lower thresholds internally calculated from a
specified reference value and tolerance
About comparisons based on a reference value and tolerance
When the reference value and tolerance method is selected, thresholds are calculated as
follows:
Upper threshold = reference value × (100 + tolerance [%]) / 100
Lower threshold = reference value × (100 - tolerance [%]) / 100
Measured values are displayed as a percentage relative to the reference value, calcu-
lated as follows:
Relative value = (measured value - reference value) / reference value× 100 [%]