EasyManua.ls Logo

Hioki dt4212 - Electrical Characteristics

Hioki dt4212
154 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
53
Electrical Characteristics
5.2 Electrical Characteristics
Noise
suppression
NMRR
DCV: -45 dB or more (50 Hz/60 Hz)
Noise
suppression
CMRR
DCV: -100 dB or more (DC/50 Hz/60 Hz, 1k
unbalance)
ACV: -60 dB or more (DC/50 Hz/60 Hz, 1k
unbalance) (However, -45 dB or more for the 1000 V
range)
Response time
(Auto range)
DCV: 1.2 to 1.4 seconds (0 V 100 V auto range
operation)
ACV: 0.7 to 0.9 seconds (0 V 100 V auto range
operation)
: 1.2 to 1.4 seconds (In nity 0
auto range
operation)
Display update
rate
Measured value:
3 times/s (after the range is xed, excluding the
resistance, continuity, electrostatic capacity, frequency)
2 times/s (resistance, continuity)
0.5 to 2 times/s (varies depending on the electrostatic
capacity)
5 times/s (frequency)
Dielectric
strength
Between the measurement terminal and case
7.06 kV AC sine wave (50 Hz/60 Hz, 60 seconds)
Maximum rated
voltage between
terminals
V terminal: 1000 V DC/AC or 10
6
V • Hz
Maximum rated
current between
terminals
Current terminal (A): 10 A DC/10 A AC
Current terminal (A, mA) 400 mA DC/400 mA AC
Maximum rated
voltage between
input terminals
and ground
1000 V AC (Measurement category II)
600 V AC (Measurement category III)
Anticipated transient overvoltage: 6000 V

Table of Contents