Accuracy Table
5.3 Accuracy Table
Accuracy warranty period 1 year
Regulated power supply range
2.4 V ± 0.15 V or more (until lights
up)
Accuracy guarantee for
temperature and humidity
23°C ± 5°C (73°F ± 9°F), 80%RH or less
(non-condensating)
Temperature characteristic Adds “Measurement accuracy × 0.1/°C”
(excluding 23°C ± 5°C (73°F ± 9°F)).
• rdg. (reading or displayed value): The value currently being measured and
displayed on the measuring instrument.
• dgt. (resolution): The smallest displayable unit, i.e., the input value that
causes the digital display to show a “1” as the least-signifi cant digit.
1
AC voltage
Range
Accuracy
*1
Input impedance
40 to 500 Hz
400.0 mV
*2
±1.0% rdg. ±10 dgt.
11M
±2%
100 pF or less
4.000 V ±1.0% rdg. ±5 dgt.
40.00 V
±1.0% rdg. ±5 dgt.
10M
+2%
100 pF or less
400.0 V
1000 V
•
Overload protection: 1100 V DC/1100 V AC or 10
6
V • Hz (energized for 1 minute)
Transient overvoltage: 6000 V
• Crest factor (Only the DT4212 (True RMS measurement model)): The crest
factor is 2 up to 2800 counts and reduces linearly to 1.5 at 4000 counts.
*1: The accuracy is specifi ed in 1% or more of the range.
*2: Only the manual range.