EasyManua.ls Logo

Hioki dt4212 - Page 69

Hioki dt4212
154 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
65
Troubleshooting
Symptom Check and/or remedy
The measurement
value does not
appear.
Even after the
measurement, 0
(zero) still appears.
Even after short
circuit of the probe,
the measured value
does not appear.
Zero adjustment is
not possible.
If the measured current value does not appear,
check that the fuse is not blown.
Check method: “Check that the fuse is not broken.”
(p. 32)
If the fuse is blown, replace it with the speci ed
fuse. (p. 67)
If the measured current value does not appear,
check that the fuse holder is not deformed.
When removing the fuse, the holder is deformed
if excessive force is applied. Pinch it with needle-
nose pliers and restore the shape of the fuse
holder.
Check that the test lead is not broken.
Perform the continuity check to con rm the
continuity of the test leads. (p. 30)
If the test lead is broken, replace the lead.
Check that the test leads have been inserted at
the ends.
Check that the measurement method is correct.
If no problems have been found, the instrument
may be malfunctioning. Send the instrument for
repair.
The display does not
stabilize and the value
uctuates; it is dif cult
to read the value.
Check that the input signal is within the input range
for the instrument.
Other inquiries
Question Solution
Would like to perform
zero adjustment.
Using the relative value display function, zero
adjustment can be performed. (p. 48)

Table of Contents

Related product manuals