22
WARNING
To avoid an electric shock, observe the following
precautions:
Prior to measurement, please make sure that the
earthing electrode has been disconnected from
the distribution system. The measurement cables
L9841, L9842-11, L9842-22, L9843-51, and L9843-52
are measurement cables with the maximum rated
voltage of 50 V (between input terminals and the
ground) and are designed to measure the earth
resistance of an earthing electrode disconnected
from the distribution system.
Turn off the instrument before connecting cables
and test leads.
Connect measurement cords or test leads to the
terminals securely. If a terminal is loose, the contact
resistance will increase, resulting in overheating,
equipment burnout, or a re.
WARNING
To prevent an electric shock, conrm that the white or
red portion (insulation layer) inside the cable of L9787
Test Lead and L9844 Measurement Cable are not
exposed. If a color inside the cable is exposed, do not
use the cable.
CAUTION
If the batteries are exhausted, the warning tone will
not sound even if a voltage is applied due to a wrong
connection. Always check the batteries before starting
to use the instrument.
To avoid damaging the cables and test leads, grasp the
connector, not the cable, to detach the cable.