Inspection Before Use
2.1 Inspection Before Use
Verify that it operates normally to ensure that no damage occurred
during storage or shipping. If you nd any damage, contact your
authorized Hioki distributor or reseller.
Check items Solution
Is the battery level sufcient? Perform a battery check to verify that the
batteries are still good.
(p. 29), (p. 36)
Is there any damage or crack in
the instrument?
Conduct visual checking.
Do not use the instrument if damage is
found. Doing so will cause an electric
shock. Send the instrument for repair.
Is there any foreign material
(sand etc.) inside the
measurement terminals?
Remove all foreign materials if any. If it
cannot be removed, the instrument needs
to be repaired.
Does the needle of the
galvanometer rest on the mark
▼?
Keeping the instrument horizontal, check
if the needle rests on the mark.
If the needle does not rest on the mark,
have the needle rest on the mark ▼ by
turning the zero adjust screw (ADJUST)
with a at-blade screwdriver without
pressing the MEASURE button.
Are the coating of the test leads
neither ripped nor torn and no
internal white parts or metal parts
exposed?
Do not use any damaged cables and
leads. Doing so may cause an electric
shock. Replace them with new ones.