2.9 PV
Ω
Measurement
Insulation resistance is a ratio of a voltage applied and
a leakage current. Indication values are not sometimes
stable depending on the object to be measured but it is
not a failure.
The MEASURE key should be fully pressed until the
hot-line warning indication lights. Insufficient press
may result in an incorrect measurement.
After using this instrument, turn OFF the rotary switch.
When you test an electrical line connected to
equipment having a withstand voltage lower than the
test voltage or equipment/part with an unknown
withstand voltage, it is recommended to make any
measurement after they are disconnected from the
line.
As solar panel to ground capacitance is larger, it may
take time until measured values become stable.
The PVΩ measurement function does not comply with
EN61557 requirements. To perform measurements in
conformity with EN61557, select the insulation
resistance range.
If open voltage of solar cell string is higher than test
voltage, accurate measurement cannot be achieved.
PVΩ500 V range should be used with an open voltage
500 V or lower and PVΩ1000 V range for an open
voltage 1000 V or lower.
If a voltage higher than test voltage is being generated,
no measurement is available.
•As PVΩ measurement uses a current-limiting resistor
of 1 MΩ connected to EARTH terminal, the output
voltage is divided by the 1 MΩ resistor and the resistor
connected between the measuring terminals. For
example, when a 10MΩ resistor is used, the output
voltage is divided by 1 MΩ and 10 MΩ.
During hours such as in the nighttime while a solar
panel is not generating power, make measurements
with P-N shorted.