*As the PVΩ function employs a 1 MΩ current-limiting resistor at the EARTH terminal,
the output voltage is divided by the 1 MΩ resistor and a resistor connected between
measuring terminals.
For example for an open voltage measurement using a DMM with its input imped-
ance at 10 MΩ, the output voltage is divided by the 1 MΩ and 10 MΩ.
Response time
With in 4.0 s
(Starting measurement to display)
HIOKI IR4051A981-09