TLi
IQ
System User Manual 4 29
Section 4 Software Functions - Detailed Descriptions
CHANGE SETUP TLi
IQ
QCette
®
SETUP
Option 6 on the Main Menu screen allows the user to change the
Date/Time, Autoprint, or Incubation Mode, or to perform QCette SETUP
from the SETUP MENU.
The SETUP MENU is displayed over two screens. Each screen can be
accessed by using and keys.
The QCette SETUP initializes the QCette for use in evaluating the
performance of the analyzer. During the initialization process, the
performance criteria of the analyzer are established. QCette SETUP
must be performed PRIOR to running the QCette as a quality control
device.
Select option 4 on the SETUP MENU to begin QCette SETUP.
fFN MAIN MENU
4-ACCESS DATA
5-VIEW SETUP
6-CHANGE SETUP ↑↓
SETUP MENU
1-DATE/TIME
2-AUTOPRINT
3-INCUBATION MODE
SETUP MENU
4-QCette SETUP