51
Scale Calibration Procedure (lbs. or kg) with Remote Display*
4. Reinstall the platter onto the scanner/Diva scale.
Figure 44. Install Platter
5. Go to Group 1 Step 8 (Weight Unit) of the Service Mode by pressing the Zero ( ) key several times until
‘Group 1 Step 8’ is displayed on the Remote Scale Display screen (see Figure 45 below).
x Depending on the current Weight Unit setting, this value will either be 1, 2, 3, or 4.
Figure 45. Remote Scale Display Group 1 Step 8 example shown
6. Verify the weight unit setting. Verify the Group 1 Step 8 setting using the table shown below. Press the
Zero ( ) key if correct. If the lb (Pounds) or kg (Kilograms) basic range setting needs to be changed,
press the Tare ( ) key to scroll through the selectable options. Press the Zero ( ) key to accept the
selected weight unit range.
Group 1 Step 8 (x)
Selectable Weight Unit Ranges
1
lbs range (0 to 30 lbs) **
2
kg range (0 to 15kg)
3
kg multi-interval (0 to 6kg, 6 to 15kg) **
4
lbs multi-interval (0 to 15 lbs, 15 to 30 lbs)
** Common ranges used as default setting.
*These procedures are for MS2420/MS2430 Scanner/Diva Scale models only.