4.6.1 XNX Electrochemical (EC) Sensor Installation
CAUTION
For biased sensors (e.g. Nitrogen Dioxide) remove the sensor stabilizer from the bottom of the
sensor prior to installation.
UsingFigure11asaguide,followtheprocedurebelow:
1. Checkthatthelabelonthenewsensoristhecorrectgastype.
2. Unscrewtheweatherproofcover,loosentheretainerlockingscrew
withthesuppliedhexkeyandunscrewthesensorretainer.
3. Pluginthenewsensortakingcaretoalignthesensorpinswiththe
connector.
4. Retthesensorretainer,tightenthelockingscrewwiththesupplied
hexkeyandrettheweatherproofcover.
5. Countdowntimeofupto180seconds(dependentonsensortype)is
displayed.
6. Acknowledgementofthegastypewillberequiredbefore
proceeding.Formoreinformationonsettinggastype,seetheXNX
TechnicalManualSection2.51GasSelection.
7. Afterthesensorisinstalledandthegastypeisconrmed,the
Range,alarmlevelsandotherimportantsettingsmustbeset;see
appropriatesectioninSection6-PoweringtheXNXfortheFirst
Time.
8. OncetheXNXhasbeencongured,calibratethedetectorfollowing
theproceduresinSection8.1-Calibration.
Transmitter
Sensor Retainer &
Locking Screw
2
1
3
4
5
New Sensor
Weatherproof Cap
Figure 11. Installing the plug-in sensor