EasyManua.ls Logo

Honeywell XNX - Page 54

Honeywell XNX
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
XNX Universal Transmitter Quick Start Guide
54
Sensor
calibrated
to detect
Sensor used to detect
8* 7* 6* 5* 4* 3* 2* 1*
8* 1.00 1.24 1.52 1.89 2.37 2.98 3.78 4.83
7* 0.81 1.00 1.23 1.53 1.92 2.40 3.05 3.90
6* 0.66 0.81 1.00 1.24 1.56 1.96 2.49 3.17
5* 0.53 0.66 0.80 1.00 1.25 1.58 2.00 2.55
4* 0.42 0.52 0.64 0.80 1.00 1.26 1.60 2.03
3* 0.34 0.42 0.51 0.64 0.80 1.00 1.27 1.62
2* 0.26 0.33 0.40 0.50 0.63 0.79 1.00 1.28
1* 0.21 0.26 0.32 0.39 0.49 0.62 0.78 1.00
NOTE
Since combustible sensors require oxygen for correct operation, a
mixture of gas in air should be used for calibration purposes. Assuming
average performance of the sensor, the sensitivity information in Tables 1
to 3 is normally accurate to +20%.
EXAMPLE
If target gas to be detected is Butane and the calibration gas available is Methane
(50% LEL):
1. Lookupthestarratingforeachgasinthersttableonpage51:
Butane4*andMethane6*.
2. Checkthemetersettingsfor50%LELcalibrationgasinthesecond
table:78.
3. Themetershouldthereforebesetto78%togiveanaccurate
readingforButaneusing50%LELMethaneasacalibrationgas.
NOTE
It is important to calibrate the sensor at the approximate alarm levels to
allow for non-linearity of the sensors at gas concentrations above 80% LEL.

Table of Contents

Other manuals for Honeywell XNX

Related product manuals