Table
8-9.
H.V. Power Supply Voltage Measurement
Conditions
1. Set
controls as follows:
LINE
power
ON.
Plug
ins not installed.
No signal
input
I NT
fully
CCW
SCALE
fylly
CCW
FOCUS
fully
CCW
2. All
voltages are
referenced
to ground.
3. Low voltages
measured with HP
Model
414A
Auto Voltmeter
(100 input
impedance).
4. To measure
high voltages
use HP
Model
K05
3440A 1000:1
Divider Probe
and HP Model
3440A
Digital Voltmeter
with
HP Model
3441A or
3444A plug-in.