Model
3465B
Section
III
3-21.
CURRENT
MEASUREMENTS.
rWWWYVWVJ
A
f
CAUTIONS
Do
not
exceed
a
maximum
input
voltage
of
250
V
dc
+
peak
ac
or
a
maximum
dc
or
ac
rms
input
current
of
2
A
or
the
amps
fuse,
located
directly
behind
the
"A"
terminal,
will
open.
See
the
following
paragraph
for
replace
ment
instructions.
3-22.
The
Multimeter
is
protected
from
the
application
of
excessive
current
by
a
2
A
fuse
located
directly
behind
the
front
panel
"A"
terminal.
If
it
is
necessary
to
replace
this
fuse,
use
the
side
slots
on
the
"A"
terminal
to
rotate
the
terminal.
The
terminal
and
fuse
will
protrude
from
the
front
panel.
Remove
the
terminal
and
fuse,
replace
fuse
with
a
2
A
rated
fuse
as
listed
in
Table
6-3
Miscellaneous
Parts
General,
and
designated
Fl.
3-23.
AC
Current
Ranges.
3-24.
AC
current
measurements
are
specified
over
a
fre
quency
range
of
40
Hz
to
20
kHz.
There
are
five
current
ranges
from
200
ph
to
2000
mA.
See
current
measure
ments
Caution
in
Paragraph
3-21.
3-25.
DC
Current
Ranges.
3-26.
DC
Current
measurements
can
be
made
on
five
cur
rent
ranges
from
200
pA
to
2000
mA.
See
current
measure
ments
caution
in
Paragraph
3-21.
3-27.
OHMS
MEASUREMENTS.
A
Do
not
apply
voltage
greater
than
±
250
V
dc
-r
Peak
AC
between
the
ohms
and
common
input
terminals
in
the
ohms
function
or
damage
to
the
instrument
will
occur.
3-28.
Ohmmeter
Ranges.
3-29.
Resistance
measurements
can
be
made
on
six
ranges
from
200
ohms
to
20
megohms.
Both
input
terminals
(f2
and
COM)
are
floating
with
respect
to
circuit
ground.
3-30.
Ohmmeter
Reference
Current.
3-31.
The
ohmmeter
reference
current
through
the
un
known
resistance
for
each
range
is
shown
in
Table
3-1.
Table
3-1.
Ohmmeter
Current
Through
Unknown.
Current
Range
Through
Unknown
200
a
1
mA
2
ka
1
mA
20
ka
10
mA
200
kn
10
mA
2000
kn
1
mA
20
Mn
0.1
mA
Maximum
open-circuit
voltage
at
the
ohms
input
terminals
is
less
than
5
V.
3-3/3-4