PERFORMANCE
TESTS
EQUIPMENT:
DUTBoxX.........ocuunt
HP
16361A/16362A
Test
Leads...............
HP
P/N
16361-61605
Note
User
built
test
fixture/leads
or
DUT
box
must
take
residual
impedance
into
consideration.
PROCEDURE:
1.
Connect
Test
Leads
(HP
P/N
16361-61605)
between
4262A
UNKNOWN
termi-
nals
and
HP
16361A
DUT
Box
(see
Figure
4-5).
When
TEST
SIGNAL
frequency
is
10kHz,
use
HP
16362A
in
place
of
HP
16361A.
2.
Set
4262A
controls
as
follows:
DCBIAS.
...
.
e
OFF
FUNCTION.
.
...
i
e
L
LOSS.
.
e
D
LCRRANGE
.............cvnan.
AUTO
D/IQRANGE.
...........ciivin..
AUTO
TRIGGER.
.
.........c
i,
INT
3.
Set
HP
16361A/16362A
LCR
RANGE
to
100mH.
4.
Confirm
that
L
accuracy
is
within
the
test
limits
shown
in
table
below:
Note
Test
limits
below
are
given
for
100mH
in-
ductance
measurement.
If
another
inductance
value
is
measured,
refer
to
SPECIFICATIONS
above.
E
.
T
FS\;I;
s'IG
C{\?ggéT
TEST
Limits
L
Readout
PRL
Calibrated
Value
+
3
counts
120Hz
SER
Calibrated
Value
+
4
counts
PRL
Calibrated
Value
+
5
counts
1kHz
SER
Calibrated
Value
+ 4
counts
PRL
Calibrated
Value
t
5
counts
10kHz
SER
Calibrated
Value
+ 4
counts
Note
Error
caused
by
stability
of
standard
component
is
not
taken
into
account
for
test
limits
in
table
above.
If
this
test
fails,
refer
to
Section
VIII,Service.