Chapter 2  Operating Your Universal Counter
Using the Save and Recall Menus
2-54 Operating Guide
2
7 Press Enter key to complete the 1.5V entry.
8 Press 50
Ω
/1M
Ω
 key to set input impedance to 50
Ω
.
9 Press DC/AC key to set input coupling to DC.
10 To save these settings, which make up your measurement setup, simply 
perform the following steps:
a. Press Save & Print key.
SAVE:    NO
 is displayed.
b. Press the appropriate arrow key until 
SAVE:     1
 is displayed.
c. Press Enter key.
The measurement setup is now saved in register 1. To recall this 
measurement setup, perform the recall procedure in the following 
section.
To Use the Recall Function
This procedure describes how to RECALL the measurement setup you’ve 
saved in the previous procedure. 
NOTE
You cannot save a setup to register 0 (RECALL    0). The Counter 
automatically saves the current setup to register 0 before executing a 
recall function. You can, however, recall the setup in the register 0 if you 
need to, which is the purpose of this “backup” storage register.
1 Cycle POWER key to preset the Counter.
Note the Counter is set to measure the frequency of a signal applied to 
channel 1 as indicated by the Freq and Ch1 annunciators.
2 Press Recall key until 
RECALL   1
 is displayed, and press Enter key or 
let the Counter time out.
Note that the measurement setup has been recalled as indicated by the 
Period and Ch1 annunciators, and 50Ω and DC LEDs.