Chapter 2  Operating Your Universal Counter
Using the MEASURE Menu Keys
Operating Guide 2-11
2
To Measure Frequency
1 Connect power source to Counter, and turn on Counter.
All segments of the front-panel display will light up while the Counter 
performs its power-on self-test, and then dashes are displayed. 
The Counter is now ready to measure frequency of a signal applied to 
CHANNEL 1 input. Note that the Freq and Ch1 annunciators light. 
2 Connect an input signal to CHANNEL 1.
The Counter should automatically display the measured frequency of the 
input signal. The trigger levels are set to the 50% points of the signal. 
However, you may disable auto triggering, and change the trigger levels 
and slopes.
To set up CHANNEL 1’s coupling, impedance, and triggering conditions to 
match the input signal you are trying to measure, refer to the procedures 
“To Select Input Coupling and Impedance” and “To Set Input Channel 
Trigger Level/Sensitivity” in Chapter 1, “Getting Started,” or the section 
titled “Using CHANNEL 1 and CHANNEL 2 Input Conditioning Keys” in 
this chapter.
3 Press the Freq/Ratio key until 
FREQUENCY  2
 is displayed to measure 
the frequency of an input signal applied to CHANNEL 2.
FREQUENCY  2
 is momentarily displayed, the Freq and Ch2 
annunciators light, and the Counter is ready to measure frequency of a 
signal applied to CHANNEL 2 input. The trigger levels are set to the 
50% points of the signal. However, you may disable auto triggering, and 
change the trigger levels and slopes.
4 To measure the frequency of a signal applied to optional CHANNEL 3 
input, repeat step 3 until 
FREQUENCY  3
 is displayed.
FREQUENCY  3
 will be momentarily displayed and the Freq and Ch3 
annunciators will light. The Counter will be ready to measure the 
frequency of a signal applied to CHANNEL 3 input.
Note that the Frequency 3 menu item is not available in Counters that do 
not contain optional Channel 3.