Chapter 2  Operating Your Universal Counter
Using the Utility Menu
2-58 Operating Guide
2
To Set the HP-IB Address
Each device on the HP-IB must have a unique address. You can set the 
Counter’s address to any value between 0 and 30. The address is set to “3” 
when the Counter is shipped from the factory.
Selecting Operating Mode (Talk/Listen, Talk-Only)
The Counter has two HP-IB operating modes:
• Addressed (talk/listen)—This mode is for bi-directional 
communication. The Counter can receive commands and setups from 
the computer, and can send data and measurement results.
To select the talk/listen operating mode, set the Counter’s 
HP-IB address from 0 to 30. Refer to the following sub-section titled 
“Setting the HP-IB Address” for instructions on how to select an 
HP-IB address from the front panel.
• Talk-only—In this mode, the Counter can send data to a printer. 
It cannot receive commands or setups from the computer. 
To select the talk-only operating mode, set the Counter’s 
HP-IB mode to “TALK”. Refer to the section titled “To Select the 
HP-IB Talk-Only Mode for Printing” for instructions on how to select 
talk-only operating mode.
Setting the HP-IB Address
1 Press and hold Recall (Utility) key, then cycle POWER key.
REV:  n
 is displayed; where “n” represents the numeric firmware revision 
code.
2 Press Recall (Utility) key until 
HP-IB:
 is displayed.
HP-IB:        3
 is displayed, for example.
3 To set the address to 15 (for example), use the appropriate arrow keys to 
enter the value 15, then press Enter key.
NOTE
BE SURE to press the Enter key to complete the numeric entry.
The address is now stored in non-volatile memory, and does not change 