Chapter 2 Operating Your Universal Counter
Using the Utility Menu
Operating Guide 2-61
2
Example Procedure for Running the Self Test
1 Press and hold Recall (Utility) key, then cycle POWER key.
2 Press Recall (Utility) key until
TEST:
is displayed.
3 Press any one of the arrow keys to select the desired menu choice in the
menu item (for example, the
ALL?
menu choice).
4 Press Enter key to start the test.
5 If you are asked to do something, press Enter key when you have
completed the requested task.
A pass or fail message will be displayed.
To Configure the RS-232 Serial Port for Printing
Software pacing, baud rate, parity, data bits, and stop bits parameters for
each port are user-selectable and independent of configuration of the
other.
The table below lists the configuration factory-default values for the
RS-232 port.
1
When parity is EVEN or ODD, data bits are fixed at 7. When parity is NONE, data bits are fixed at 8.
Parameter Default Possible Choices
Software Pacing XON XON or NONE
Baud Rate 9600 300, 1200, 2400, 9600, or 19200
Parity NONE EVEN or ODD
1
NONE
Data Bits 8 FIXED at 7 FIXED at 8
Stop Bits 1 FIXED
Full Duplex ON ON or OFF
DTR HIGH PACE, HIGH, LIMIT