Figure
1
OVercompensation
causes
pulse
peaking.
Figure
2
Correct
compensation
with
a
flat
pulse
top.
Figure
3
Undercompensation
causes
pulse
rolloff.
The
Oscilloscope
at
a
Glance
To
co11ect
a
s1g111
to
tile
oscilloscope
You
should compensate
10:1
probes to match their characteristics to the
oscilloscope. A poorly compensated probe can introduce measurement
errors.
To
compensate a probe, follow these steps.
1 Connect the
10:1
probe from channel 1 to the front-panel probe
adjust signal on the oscilloscope.
2 Press I
Autoscale
I .
3
Use
a nonmetallic tool to adjust the trimmer capacitor on the
probe
for
the flattest pulse possible as displayed on the
oscilloscope.