1370
1390,
1400
Read
the
value.
Return
the
instrument
to
local
state,
and
leave
the
subroutine.
Setting
the
Filter
The
Alter
command
on
the
HP
8152A
selects
between
a
2Hz
(filter
ON)
and
an
8Hz
(Filter
OFF)
low
pass
filter.
This
filter
is
not
available
on
the
HP
8153A.
The
filter
command
in
HP
8152A
Compatability
mode
selects
between
an
averaging
time
of
500ms
(filter
ON)
and
an
averaging
time
of
200ms
(filter
OFF).
Switching
the
instrument
into
HP
8162A
Compatability
mode,
does
not
alter
the
averaging
time.
You
should
use
*RST,
or
a
filter
ON
or
OFF
command
to
set
the
averaging
time
to
500ms
or
200ms.
H-4
Remote
Operation