Running
the
Loss
Application
After
pressing
[lossI
the
instrument
checks
to
see
if
both
a
source
and
sensor
module
are
present.
If
one
or
both
are
missing
the
character
field
shows
the
message
CONFIG
?.
The
instrument
does
not
enter
the
loss
application.
If
both
modules
are
present:
The
sensor
wavelength
is
automatically
set
to
the
wavelength
of
the
source.
(When
you
stop
the
application
running,
the
parameters
for
the
sensor
are
automatically
reset
to
the
values
in
use
before
you
started
the
application.)
The
operation-indicator
shows
APPL,
this
stays
lit
while
the
instrument
is
in
the
loss
application.
The
instrument
shows
LOSS
in
the
character
field.
The
instrument
reads
the
reference
level
when
you
start
the
application
running.
To
read
the
input
signal
and
use
it
as
the
reference,
do
the
following:
1.
Make
sure
that
the
device
under
test
is
not
in
the
system.
2.
Make
sure
that
the
source
is
not
active.
The
source
is
active
whenever
the
green
LED
on
its
front
panel
lights.
3.
Connect
the
source
to
the
input
of
the
system
without
the
device
under
test.
4.
Connect
the
output
of
the
system
to
the
sensor.
Figure
3-1.
Measuring
the
Reference
for
the
Loss
Application
5.
P
ress
The
instrument
takes
the
reference
and
then
runs
the
application.
While
the
application
is
running,
the
RUN
operation-indicator
lights.
Local
Control
Menu
Mode
3-3