3-4
Front/Rear Panel
Front Panel Features
10. INSTRUMENT STATE function block. These keys allow you to control
channel-independent system functions such as the following:
• copying, save/recall, and HP-IB controller mode
• limit testing
• tuned receiver mode
• frequency offset mode (Option 089)
• test sequence function
• time domain transform (Option 010)
HP-IB STATUS indicators are also included in this block.
11. key. This key returns the instrument to either a known factory
preset state, or a user preset state that can be defined. Refer to Chapter 8 ,
“Preset State and Memory Allocation” for a complete listing of the
instrument preset condition.
12. (ES) R CHANNEL connectors. These connectors allow you to apply an input
signal to the analyzer's R channel, for frequency offset mode.
13. (ES) PORT 1 and PORT 2. These ports output a signal from the source and
receive input signals from a device under test. PORT 1 allows you to
measure S
12
and S
11
. PORT 2 allows you to measure S
21
and S
22
.
13. (ET) REFLECTION and TRANSMISSION ports. The REFLECTION port
outputs a signal from the source and allows you to make reflection
measurements. The TRANSMISSION port receives input signals from a
device under test and allows you to make transmission measurements.
Preset