6-10
Error Messages
Error Messages in Alphabetical Order
FORMAT NOT VALID FOR MEASUREMENT
Error Number
75
Conversion measurements (Z or Y reflection and transmission) are not valid
with Smith chart and SWR formats.
FORMATTING DATA
Information
Message
The list information is being processed for a list data output to a copy device and
stored in the copy spool buffer. During this time, the analyzer's resources are
dedicated to this task (which takes less than a few seconds).
FREQ OFFSET ONLY VALID IN NETWORK ANALYZER MODE (ES Option 089 Only)
Error Number
140
You can only make frequency offset measurements in the network analyzer
mode.
FREQS CANNOT BE CHANGED, TOO MANY POINTS
Error Number
204
The number of points selected for setting the low pass transform frequencies is
too high. Reduce the number of points so that the low pass criteria is met.
FUNCTION NOT AVAILABLE
Error Number
202
The function you requested over HP-IB is not available on the current
instrument.
FUNCTION NOT VALID
Error Number
14
The function you requested is incompatible with the current instrument state.
FUNCTION NOT VALID DURING MOD SEQUENCE
Error Number
131
You cannot perform sequencing operations while a sequence is being modified.
FUNCTION NOT VALID FOR INTERNAL MEMORY
Error Number
201
The function you selected only works with disk files.
FUNCTION ONLY VALID DURING MOD SEQUENCE
Error Number
163
You can only usethe capability when you are building a
sequence. Attempting to use this softkey at any other time returns an error
message and no action is taken.
GOSUB SEQUENCE