6-9
Error Messages
Error Messages in Alphabetical Order
DISK READ/WRITE ERROR
Error Number
189
There may be a problem with your disk. Try a new floppy disk. If a new floppy
disk does not eliminate the error, suspect hardware problems.
DISK WEAR - REPLACE DISK SOON
Error Number
49
Cumulative use of the disk is approaching the maximum. Copy files as
necessary using an external controller. If no controller is available, load
instrument states from the old disk and store them to a newly initialized disk
using the save/recall features of the analyzer. Discard the old disk.
DOMAIN RESET
Error Number
67
Time domain calculations were reset due to a change in the frequency range,
sweep type, or number of points. Perform a new time domain transform on the
new state.
DUPLICATING TO THIS SEQUENCE NOT ALLOWED
Error Number
125
A sequence cannot be duplicated to itself.
EXCEEDED 7 STANDARDS PER CLASS
Error Number
72
When modifying calibration kits, you can define a maximum of seven standards
for any class.
FILE NOT COMPATIBLE WITH INSTRUMENT
Information
Message
You cannot recall user graphics that had been saved on an earlier model of
analyzer with a monochrome display. These files cannot be used with the
analyzer.
FILE NOT FOUND
Error Number
192
The requested file was not found on the current disk medium.
FILE NOT FOUND OR WRONG TYPE
Error Number
197
During a resave operation, either the file was not found or the type of file was
not an instrument state file.
FIRST CHARACTER MUST BE A LETTER
Error Number
42
The first character of a disk file title or an internal save register title must be an
alpha character.