6-19
Error Messages
Error Messages in Alphabetical Order
REQUESTED DATA NOT CURRENTLY AVAILABLE
Error Number
30
The analyzer does not currently contain the data you have requested. For
example, this condition occurs when you request error term arrays and no
calibration is active.
SAVE FAILED. INSUFFICIENT MEMORY
Error Number
151
You cannot store an instrument state in an internal register due to insufficient
memory. Increase the available memory by clearing one or more save/recall
registers and pressing , or by storing files to a disk.
SEGMENT #n POWER OUTSIDE RANGE LIMIT
Information
Message
The selected power range does not support the power level of one or more
segments in the swept list table. This message appears when swept list mode is
not on and reports the first segment that is out of range. Change the segment
power or change the power range.
SEGMENT #n START FREQ OVERLAPS PREVIOUS SEGMENT
Information
Message
A segment entered in the swept list table caused one or more frequency
segments to overlap. This message appears when swept list mode is not on and
reports the first segment that is overlapping another. Change the frequency
ranges of the overlapping segments.
SELECTED SEQUENCE IS EMPTY
Error Number
124
The sequence you attempted to run does not contain instrument commands.
SELF TEST #n FAILED
Service Error
Number 112
Internal test #n has failed. Several internal test routines are executed at
instrument preset. The analyzer reports the first failure detected. Refer to the
service guide for troubleshooting information on internal tests and the
self-diagnose feature.
SEQUENCE ABORTED
Error Number
157
The sequence running was stopped prematurely when you pressed the
key.
SEQUENCE MAY HAVE CHANGED, CAN'T CONTINUE
Error Number
153
When you pause a sequence, you cannot continue it if you have modified it. You
must start the sequence again.