HP4402.book : chapter2.fm 69 Thu Feb 11 15:53:59 1999
Power Meter Operation
Self Test
HP E4418B User’s Guide 2-69
Remote Testing
To invoke the remote self test, the IEEE 488.1 compliant standard
command, *TST? is used. This command runs a full self test and returns
one of the following codes:
0 - no tests failed
1 - one or more tests failed
The remote self test consists of the following tests:
ROM checksum
RAM
RAM Battery
Display Assembly
Calibrator
Measurement Assembly
Communications Assembly (Implicit)
The communications assembly is tested implicitly, in that the command
will not be accepted or return a result unless the HP-IB interface is
functioning correctly.
Refer to “Test Descriptions”, on page 2-70 if you require a description of
each individual test.
When the *TST? command is executed, the screen is cleared. As each test
takes place, the name of the test is listed on the screen. While a test is
running, the message “Testing...” appears next to the name of the test.
As each stage of the test is completed, the message “Testing...” is
replaced by either the message “Passed” or “Failed”.

Table of Contents

Other manuals for HP E4418B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP E4418B and is the answer not in the manual?

HP E4418B Specifications

General IconGeneral
BrandHP
ModelE4418B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals