HP4402.book : chapter4.fm 10 Thu Feb 11 15:53:59 1999
Error Messages
Error Messages
4-10 HP E4418B User’s Guide
-330 Self-test Failed;
The -330, “Self-test Failed” errors indicate that you have a problem
with your power meter. Refer to “Contacting Hewlett-Packard”, on
page 2-74 for details of what to do with your faulty power meter.
-330 Self-test Failed;Measurement Channel Fault
Refer to “Measurement Assembly”, on page 2-70 if you require a
description of the Measurement Assembly test.
-330 Self-test Failed;Option 001 Battery requires replacement
The option 001 battery is not charging to a satisfactory level and should
be replaced.
-330 Self-test Failed;RAM Battery Fault
Refer to “RAM Battery”, on page 2-70 if you require a description of the
battery test.
-330 Self-test Failed;Calibrator Fault
Refer to “Calibrator”, on page 2-71 if you require a description of the
calibrator test.
-330 Self-test Failed;ROM Check Failed
Refer to “ROM Checksum”, on page 2-70 if you require a description of
the ROM Checksum test.
-330 Self-test Failed;RAM Check Failed
Refer to “RAM”, on page 2-70 if you require a description of the RAM
test.
-330 Self-test Failed;Display Assy. Fault
Refer to “Display”, on page 2-71 if you require a description of the
Display test.
-330 Self-test Failed;Confidence Check Fault
Refer to “Confidence Check”, on page 2-68 if you require a description of
this test.
-330 Self-test Failed;Serial Interface Fault
Refer to “Serial Interface”, on page 2-70 if you require a description of
this test.
-350 Queue overflow
The error queue is full and another error has occurred which could not
be recorded.