HP4402.book : chapter4.fm 8 Fri Feb 12 15:32:45 1999
Error Messages
Error Messages
4-8 HP E4419B User’s Guide
-230 Data corrupt or stale;Please zero and calibrate Channel A
When CAL[1|2]:RCAL is set to ON and the sensor currently connected
to channel A has not been zeroed and calibrated, then any command
which would normally return a measurement result (for example
FETC?, READ?, or MEAS?) will generate this error message.
-230 Data corrupt or stale;Please zero and calibrate Channel B
When CAL[1|2]:RCAL is set to ON and the sensor currently connected
to channel B has not been zeroed and calibrated, then any command
which would normally return a measurement result (for example
FETC?, READ?, or MEAS?) will generate this error message.
-230 Data corrupt or stale;Please zero Channel A
When CAL[1|2]:RCAL is set to ON and the sensor currently connected
to channel A has not been zeroed, then any command which would
normally return a measurement result (for example FETC?, READ?, or
MEAS?) will generate this error message.
-230 Data corrupt or stale;Please zero Channel B
When CAL[1|2]:RCAL is set to ON and the sensor currently connected
to channel B has not been zeroed, then any command which would
normally return a measurement result (for example FETC?, READ?, or
MEAS?) will generate this error message
-230 Data corrupt or stale;Please calibrate Channel A
When CAL[1|2]:RCAL is set to ON and the sensor currently connected
to channel B has not been calibrated, then any command which would
normally return a measurement result (for example FETC?, READ?, or
MEAS?) will generate this error message
-230 Data corrupt or stale;Please calibrate Channel B
When CAL[1|2]:RCAL is set to ON and the sensor currently connected
to channel B has not been calibrated, then any command which would
normally return a measurement result (for example FETC?, READ?, or
MEAS?) will generate this error message
-231 Data questionable;CAL ERROR ChA
Power meter calibration failed on channel A. The most likely cause is
attempting to calibrate without applying a 1 mW power to the power
sensor.
Get other manuals https://www.bkmanuals.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP E4419B and is the answer not in the manual?

HP E4419B Specifications

General IconGeneral
BrandHP
ModelE4419B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals