SR05-D1A3-PV manual v1801 49/83
Table 6.4.2 Serial communication setting options of register 0x1001
SETTING OPTIONS
SETTING
NUMBER
BAUD RATE DATABITS STOPBITS PARITY
1
9600
8
1
none
2
9600
8
1
even
3
9600
8
1
odd
4 19200 8 1 none
5 ( = default)
19200
8
1
even
6 19200 8 1 odd
7 38400 8 1 none
8 38400 8 1 even
9 38400 8 1 odd
10 115200 8 1 none
11 115200 8 1 even
12 115200 8 1 odd
13 1200 8 1 none
14 1200 8 1 even
15
1200
8
1
odd
16 2400 8 1 none
17 2400 8 1 even
18 2400 8 1 odd
19 9600 8 2 none
22 19200 8 2 none
25
38400
8
2
none
28 115200 8 2 none
31
1200
8
2
none
34 2400 8 2 none
Register 0x1002 + 0x1003, Irradiance, provides the solar radiation output in 0.01 W/m².
The value given must be divided by 100 to get the value in W/m². MSW and LSW should
be read together in one request.
Register 0x1006, Instrument body temperature, provides the temperature of the
instrument body in 0.01 °C. The data must be divided by 100 to achieve the value in °C.
Register 0x1007, Sensor electrical resistance, sensor resistance in 0.1 . The data needs
to be divided by 10 to get the value in Ω. This register returns a 0 by default. To read the
resistance, first a measurement has to be performed. This can be done by writing 0xFF00
to coil 2. Hukseflux recommends to use this function only when necessary for diagnostics
in case of sensor failure.
Register 0x1008, Scaling factor irradiance, default scaling factor is 100