Calibration
and
Absorbance
Linearity
Each
instrument
is
calibrated
during
manufacturing
using
standards
that
are
traceable
to
the
National
Institute
for
Standards
and
Testing
(NIST),
and
is
tested
to
verify
its
linearity
to
2A.
This
preset
calibration
is
very
stable.
Absolute
calibration
can
be
verified
with
the
use
of
NIST
filters,
or
by
periodic
comparison
to
a
reference
instrument
that
is
known
to
be
calibrated
to
NIST
filters.
Calibration
may
be
confirmed
using
a
commercially
available
calibration
check
set
which
can
be
obtained
from
your
distributor.
A
periodic
verification
of
instrument
linearity
is
advised.
Since
most
lab
test
results
are
based
upon
standards
rather
than
upon
absolute
absorbances,
the
linearity
of
the
instrument
is
the
more
critical
indicator
of
instrament
performance.
CAUTION
DO
NOT
ALTER
ANY
POTENTIOMETER
SETTINGS
!
Changing
these
settings
will
make
the
factory
calibration
data
invalid.
In
the
unlikely
event
the
calibration
data
is
lost
or
corrupted,
the
absorbance
factor
is
set
to
1.000
and
the
temperature
offset
adjustments
for the
block
and
cell
are
set
to
0,0.
Do
not
enter
values
other
than
those
recorded
on
the
calibra-
tion
label
unless
absolutely necessary.