-6-
[Key sequence]
Æ
ALT + MENU
Æ
ALT + START
Æ
ALT + TEST
2.2.2 Post aging test procedures for the Refractor Body
Check the PD adjustment function
Press PD button.
Check if the PD-check lenses are inserted properly.
Turn the dial clockwise or counter-clockwise several times.
Press PD button again to finish this step.
Check the convergence function
Press F/N button several times.
Check if the Refractor Body is converging correctly.
Finish this step with the Refractor in the FAR mode for the remaining steps.
Check the initialization sequences again
Turn off the power.
Turn on the power again.
Check if the disk plates in the Refractor Body are aligned properly after the
initialization sequences.
Repeat this step 3 to 5 times.
Check the auxiliary lenses
To enter the lens-checking mode, press the following key sequence.
ALT + MENU
ALT + START
ALT + OPEN/CLOSE
Change the selection of eye to binocular:
Æ Press the OU button.
Skip SPH lenses by turning the dial clock-wise or press NEXT button.
Check the auxiliary lenses one by one.
Æ Compare the lenses in the test windows with the information on the LCD of the