EXCELON
CPE-4000 Maintenance
Ⅱ
-
21
7. Remove the Window Cover Limit Switch Ass’y.
8. Remove four screws that are connecting Feeler Ass’y and Feeler Mount Base.
9. Remove screw connecting Edging Room and Feeler Ass’y.
10. Separate the Feeler Ass’y to the right direction as below picture.
11. If there are a lot of dusts in Filter Cap, then remove them. If necessary, separate the Filter Cap.
12. Assemble the Feeler Ass’y in reverse order.
Be sure to let the Feeler Encoder Cable passes the rectangle slot of Feeler Mount Base.
13. After finishing assembling the Feeler Ass’y, check the smoothness of the Feeler Tip Ass’y.
Step 9
Step 8
Limit Switch Ass’y
Step 8
Feeler Mount Base
Step 12
Feeler Encoder Cable
Base level
Feeler Mount
Base
pin hole
Filter Cap Out
Step 11
Step 10