Display Example #28 Display Example #29
The “Set Vac Hold or Release Vac” error occurs when both the Vac Hold and Release Vac values are
(0). A valid value for one or the other (but not both) must be entered.
The “Clear Vac Hold or Release Vac” error occurs when values for both Vac Hold and Release Vac
have been entered. A valid value for one or the other (but not both) must be entered.
When all the parameters have been entered and the program is ready for saving, touch the Ok button.
By touching the Ok button, all entries are saved to the permanent memory and the display returns to
the Select Program Screen, (Display Example #30). Notice that program slot 009 is now filled with the
new program name SAMPLE 1. The new program can now be run by touching the Run button.
NOTE! By touching the Cancel button at any time during the programming, all entries will be erased
and you will be returned to the Select Program screen, (Display Example #19). Notice that program
slot 009 will still be blank.
Display Example #30
An Additional Note on Programming Vacuum
Many furnaces do not allow for the programming of a specific vacuum level, but instead provide for a
“YES / NO” or a percentage of vacuum entry, typically 90% to 100%. The Summit furnaces allow for
a more precise vacuum control. A simple rule for entering programs that do not state a specific vacuum
level in cm or inches is to enter 72 – 73 cm (28 inches). Higher vacuum levels of 74 - 75 cm (29inches)
can be programmed in; however, most vacuum pumps will have difficulty achieving these levels.
1 2
3 4
5 6
7 8
9 0
Hold Time (mm:ss)
Vac Hold (mm:ss)
Cool Time (mm:ss)
Start Vac (C)
300
Cancel Ok
00:
01:
0
0
00:
Porcelain
Low Temp (C)
Preheat (mm:ss)
Heat Rate
(C/min)
Error!
Set Vac Hold or Release Vac
Ok
1 2
3 4
5 6
7 8
9 0
Hold Time (mm:ss)
Vac Hold (mm:ss)
Cool Time (mm:ss)
Start Vac (C)
300
Cancel Ok
00:
01:
0
0
00:
Porcelain
Low Temp (C)
Preheat (mm:ss)
Heat Rate
(C/min)
Error!
Clear Vac Hold or Release Vac
Ok
08/13/12
04:15:45
009 SAMPLE 1
010 VM 13 1ST BAKE
011 VM 13 2ND BAKE
012 VM 13 GLAZE
013 VM 13 2ND GLAZE
014 MEDIUM CORRECTIVE
015 NAT ENAMEL BAKE
016 TEST 1* (Pressing Only)
Select Pro
g
ram 450 C 0cmHg
Create
Modify
Delete
Run Cancel
Doo
r