To clean the internal components (monthly or more frequently, if needed)
Note: Do not open the analyzer door while a run is in progress as it will cancel the run and discard
the sample and cartridge.
IMPORTANT: Do not use any of the following near the analyzer: organic solvents, ammonia-based
cleaning products, ink markers, sprays containing volatile liquids, insecticides, disinfectants, polish,
or room fresheners.
1. Tap the SediVue Dx
icon on the IDEXX
VetLab Station Home
screen, tap Turn off
SediVue Dx, and
then tap Shutdown.
2. Open the analyzer
door and clean the
top and bottom of
the pipetting window
with an optical tissue
dampened with
warm water. Then
dry with a dry optical
tissue.
3. Remove the cartridge
holder, clean it with
an alcohol prep pad,
dry it with an optical
tissue, and then set it
aside.
4. Clean the left side
of the pusher arm 5
times with an alcohol
prep pad (ensuring
the right edge of
the track is cleaned
thoroughly), and then
dry it with an optical
tissue.
Pusher arm
5. Press the pusher lever
to the left and repeat
step 4 on the right
side of the pusher
arm.
6a. Using a new alcohol
prep pad, clean the
optical stage and
track repeatedly to
remove all debris.
Then dry with an
optical tissue.
IMPORTANT: Do not
exert excess pressure
on the optical window.
Pusher lever
Pusher arm
Move arm
Optical
stage
Track
Maintenance

Other manuals for Idexx SediVue Dx

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Idexx SediVue Dx and is the answer not in the manual?

Idexx SediVue Dx Specifications

General IconGeneral
BrandIdexx
ModelSediVue Dx
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals