EasyManua.ls Logo

Idexx SediVue Dx - Page 26

Idexx SediVue Dx
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
26
6b. Using a new alcohol
prep pad, clean the
optical window in a
circular motion (3–4
times) and then dry
it in a circular motion
with a clean optical
tissue.
7. Remove the centrifuge shield, waste bin and drip tray. Then clean
them with warm water and dry them with an optical tissue.
8. Using a new alcohol
prep pad, clean the
move arm and then
dry it with an optical
tissue.
Move arm
9. Return the waste bin,
drip tray, centrifuge
shield, and cartridge
holder to their
locations in the
analyzer.
10. Close the analyzer
door. On the IDEXX
VetLab Station Home
screen, tap the
SediVue Dx icon
and then tap Turn
on SediVue Dx.
Maintenance
Optical
window

Other manuals for Idexx SediVue Dx

Related product manuals