- 24 -
2.5.3. Sorting components
The DE-5000 can sort components into PASS/FAIL
categories. Components may be sorted based on
resistance, capacitance, or inductance.
Note
: this function is not available in
Auto LCR
mode. Before using this function, set the meter
to either
Auto L
mode,
Auto C
mode, or
Auto R
mode.
To enter the
Sorting
mode, proceed as follows:
1.
Zero the meter. Refer to pages 14-16 for details.
2.
Press
LCR AUTO
to cycle through measurement
options until the desired impedance type is
selected.
Note: If you select Auto LCR, the sorting function
will not work.
3.
Connect one of the items that is going to be
measured to the test points.
4.
Press
SORTING
, then press
SETUP
.
Note: If you press the SORTING key while the
meter is reading outside limits (OL) or while the
reading is less than 200 counts, the Sorting function
is disabled.