Page #47 background imageLoading...
Page #47 background image
- 39 -
3.2.
Accuracy specifications
Accuracy is specified at 23°C ± 5°C, <75% RH.
All accuracy is specified as ±[(% of reading) + (value of least
significant digit)].
For most precise measurements, the meter has to be zeroed first.
Resistance:
Range Resolution 100/120Hz 1kHz 10kHz 100kHz
20.000 0.001 1.0%+3 1.0%+3 2.0%+3
200.00 0.01 1.0%+3 0.3%+2 0.3%+2 0.6%+3
2.0000 k 0.0001 k 0.3%+2 0.3%+2 0.3%+2 0.6%+3
20.000 k 0.001 k 0.3%+2 0.3%+2 0.3%+2 0.6%+3
200.00 k 0.01 k 0.5%+2 0.5%+2 0.5%+2 1.0%+3
2.0000 M 0.0001 M 1.0%+3 1.0%+3 1.0%+3
2.000M 0.001 M 2.0%+3
20.000 M 0.001 M 2.0%+3 2.0%+3
20.00 M 0.01 M 2.0%+3
200.0 M 0.1 M 2.0%+3 2.0%+3
Capacitance:
Range Resolution 100/120Hz 1kHz 10kHz 100kHz
200.00 pF 0.01 pF 1.2%+5 2.0%+5
2000.0 pF 0.1 pF 2.0%+3 0.3%+2 0.6%+3
20.000 nF 0.001 nF 2.0%+3 0.3%+2 0.3%+2 0.6%+3
200.00 nF 0.01 nF 0.3%+2 0.3%+2 0.3%+2 0.6%+3
2000.0 nF 0.1 nF 0.3%+2 0.3%+2 0.6%+2 2.0%+5
20.000 µF 0.001 µF 0.3%+2 0.6%+2 1.2%+5
20.00 µF 0.01 µF
3.0%+5
(10µF max.)
200.00 µF 0.01 µF 0.6%+2 1.0%+3
200.0 µF 0.1 µF
3.0%+5
(100µF max.)
2000.0 µF 0.1 µF 1.0%+3
2000 µF 1 µF 1.2%+3
20.00 mF 0.01 mF 1.2%+3
*If reading <2000, unit on display is pF