EasyManua.ls Logo

IFM Electronic SD8000 - 9�3�2 Settings for Quantity Monitoring Via the Preset Counter; 9�3�3 Settings for Program-Controlled Counter Reset; 9�3�4 Deactivation of the Counter Reset; 9�3�5 Counter Reset Using an External Signal

IFM Electronic SD8000
29 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
19
UK
9.3.2 Settings for quantity monitoring via the preset counter
Select [OU1] and set [ImP]�
Select [ImPS] and set the volumetric flow quantity at which output 1
switches(→9.7).
Select [ImPR] and set [no]�
> Pulse repetition is not active� The output switches ON if the value set in
[ImPS] is reached� It remains set until the counter is reset�
9.3.3 Settings for program-controlled counter reset
Select [rTo] and continue with a) or b)
a) Reset the counter manually
Press [Set] until [rES�T] is displayed, then briefly press [Mode/Enter]�
b) Enter the value for time-controlled reset
Press [Set] until the requested value is displayed (intervals from 1 hour
to 8 weeks), then briefly press [Mode/Enter]�
Press [Set] until [rES�T] is displayed, then briefly press [Mode/Enter]�
9.3.4 Deactivation of the counter reset
Select [rTo] and set [OFF]�The meter is only reset after overflow (= fac-
tory setting)�
Overflow: After the maximum value (9,999,999 Nm³) the counter is reset to
0�
9.3.5 Counter reset using an external signal
Select [OU2] and [InD]�
Select [Din2] and set the reset signal�
- [HIGH] = reset for high signal
- [LOW] = reset for low signal
- [+EDG] = reset for rising edge
- [-EDG] = reset for falling edge
LED7(→7Operatinganddisplayelements)indicatestheinputstatus
also with an active external reset�

Table of Contents