EasyManua.ls Logo

IFM PQ7809 - Output Signal Configuration; Setting Switching Limits for Outputs

IFM PQ7809
19 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
14
9.2 Setting of the output signals
9.2.1 Setting of the unit of measurement for system pressure
Select [uni] and set the unit of measurement:
[bAr], [kPa], [PSI], [inHg]�
9.2.2 Setting of the output function
Select [ou1] and set the function:
- [Hno] = hysteresis function / normally open
- [Hnc] = hysteresis function / normally closed,
- [Fno] = window function / normally open,
- [Fnc] = window function / normally closed�
Select [ou2] and set the function:
- [Hno] = hysteresis function / normally open,
- [Hnc] = hysteresis function / normally closed,
- [Fno] = window function / normally open,
- [Fnc] = window function / normally closed,
- [diA] = output 2 is used as a diagnostic output�
9.2.3 Setting of the switching limits (hysteresis function)
Make sure that the function [Hno] or [Hnc] is set for [ou1] or [ou2]�
Select [SP1] / [SP2] and set the value at which the output switches�
Select [rP1] / [rP2] and set the value at which the output switches off�
rPx is always smaller than SPx� The unit only accepts values which are
lower than the value for SPx�
9.2.4 Setting of the switching limits (window function)
Make sure that the function [Fno] or [Fnc] is set for [ou1] or [ou2]�
Select [FH1] / [FH2] and set the upper limit of the acceptable range�
Select [FL1] / [FL2] and set the lower limit of the acceptable range�
FLx is always lower than FHx� The unit only accepts values which are lower
than FHx�

Related product manuals