EasyManua.ls Logo

IFM PQ7809 - Service and Calibration Functions; Zero Point Calibration and Factory Reset

IFM PQ7809
19 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
16
9.3.4 Zero point calibration
Select [coF] and set a value between -5 % and 5 % of the measuring
span� The internal measured value "0" is shifted by this value�
As an alternative: automatic adjustment of the offset in the range 0 bar ± 5
% of the measuring span�
Make sure that the differential pressure is 0 bar or as close as possible
to the 0 bar mark�
Press [Mode/Enter] until [tcoF] appears�
Press [Set] and keep it pressed�
> The current offset value ( in %) briefly flashes�
Release [SET]�
Press [Mode/Enter] briefly (= to confirm the new offset value)�
Reset of the teached value:
Select [coF] and set the value [0]�
9.4 Service functions
9.4.1 Reading of the min/max values for system pressure
Select [Hi] or [Lo], briefly press [Set]�
[Hi] = maximum value, [Lo] = minimum value�
Delete memory:
Select [Hi] or [Lo]�
Press [Set] and keep it pressed until [----] is displayed�
Press [Mode/Enter] briefly
9.4.2 Reset of all parameters to factory setting
Select [rES]�
Press [Set] and keep it pressed until [----] is displayed�
Press [Mode/Enter] briefly
We recommend taking down your own settings in the table before carrying
out a reset (→ 13 Factory setting).

Related product manuals