• If you test I
PE
or I
functions and displayed value is close to 0.000 mA, you must repeat
the test once more after switching the L-N polarity by repetitive press and release of
either the I
PE
or I
key. The (INV) indicator lights up in such case.
If the value 2.000 M ÷ 2.600 M is displayed, the insulation resistance measurement is
correct.
If the value 0.090 mA ÷ 0.110 mA is displayed, the leakage current measurement is correct.
Contact service if the test results differ from the ones described above.