COMMAND REFERENCE
Command Reference
7/01 LDC-3700 Series 101
CHAPTER 4
The TEC:ENABle:OUTOFF? query returns the value of the status outoff enable register of the TEC
operations (things which will turn the TEC output off).
Notes The enabled TEC outoff events and conditions (which can turn the TEC output off) can be set by
using the TEC:ENABle:OUTOFF command.
Examples "TEC:ENAB:OUTOFF?" −response: 258, means that a Sensor Type Change Condition or a Voltage
Limit Condition will cause the TEC output to be turned off.
"Radix Hex; TEC:Enab:Outoff?" −response: #H17FB, means that all of the above conditions will
cause the TEC output to be turned off.
"Radix Bin; TEC:Enab:Outoff?" −response: #B1001, means that the High Temperature Limit and TE
Current Limit conditions will cause the TEC output to be turned off.
The TEC:EVEnt? query returns the value of the status event register of the TEC operations.
Notes The TEC conditions which are reported to the status byte are set via the TEC:ENABle:EVEnt
command.
The TEC event status is only cleared when the event status is read or a *CLS command is issued,
while the condition status is constantly changing.
Examples "TEC:EVE?" −response: 513, means that the Output went In or Out of Tolerance and the TE
Current Limit events have occurred since the last TEC:EVEnt? query.
"Radix Hex; TEC:Event?" −response: #H82, means that the Voltage Limit and TE Module Open
events have occurred since the last TEC:EVEnt? query.
TEC:ENABle:OUTOFF?
COMMON
DEVICE DEPENDENT
FRONT PANEL
1 − TE Current Limit Condition 256 − Sensor Change (While Output On) Event
2 − Voltage Limit Condition 512 − Output Out of Tolerance Condition
4 − N/A 1024 − Sensor Shorted (While Output On) Condition
8 − High Temperature Limit Condition 2048 − N/A
16 − TEC Interlock Changed Condition 4096 − N/A
32 − Booster Changed (While Output On) Event 8192 − N/A
64 − Sensor Open (While Output On) Condition 16384 − N/A
128 − Module Open (While Output On) Condition 32768 − N/A
TEC:EVEnt?
COMMON
DEVICE DEPENDENT
FRONT PANEL
1 − TE Current Limit 256 − Sensor Type Changed
2 − TE Voltage Limit 512 − Output Changed to be Out of Tolerance
4 − N/A 1024 − Output On/Off Changed
8 − High Temperature Limit 2048 − New Measurements Taken
16 − TEC Interlock Disabled 4096 − N/A
32 − Booster Changed 8192 − N/A
64 − Sensor Open 16384 − N/A
128 − TE Module Open 32768 − N/A