REMOTE OPERATION
Command Syntax
42 LDC-3700 Series
CHAPTER 3
For further clarity in programming, the (Boolean) values of one (1) and zero (0) may be used or their
names as indicated in the following table.
If multiple parameters are expected, they should be separated with commas. For example, to set
the Steinhart−Hart constants on the LDC−3700B Series Laser Diode Controller (C1, C2, and C3) the
following command could be sent:
TEC:CONST 1.111, 2.004, 0.456
If not all of the parameters need to be changed the other parameters may be omitted. For example:
TEC:CONST 1.111, , 0.456
A query has no space between the mnemonic and the question mark, as in:
LAS:LDI?
Table 3.1 Substitute Parameter Names
SUBSTITUTE NAME VALUE
ON 1
OFF 0
OLD 1
NEW 0
TRUE 1
FALSE 0
SET 1
RESET 0