OPERATIONS
10 LDC-3700 Series
CHAPTER 2
Note: If the instrument cannot successfully complete the power on test, an error message
of E−512 or E−513 will be displayed on the Laser Display. See Troubleshooting on page 129
for a list of error codes and actions to take when your instrument indicates these errors.
37x4B Series Default Configuration:
GPIB mode in LOCAL via front panel, or in
REMOTE via GPIB
PARAMETERS not selected
TEC and LASER adjust not selected TEC output off
TEC DISPLAY enabled, in T mode Constant T mode selected
TEC Display showing actual temperature Temperature Set Point = 20 °C
Resistance/Reference Set Point = 1 ohm or
mA or mV (depending on the setting of the
SENSOR SELECT switch)
LIM ITE set to 2.0 Amps
I
TE
Set Point = 1.0 Amp TEC STEP value = 1
LIM T
HI
set to 99.9°C GAIN = 30
TEC Tolerance values = 0.2ûC, 5 seconds C2 = 2.347 (x 10
−4
)
C1 = 1.125 (x 10
−3
) CAL PD = 0 mA/mW
C3 = 0.855 (x 10
−7
) 25 mA (3714B)
LIM I (high range) = 1000 mA (3744B), 125
mA (3724B),
12.5 mA (3714B)
LIM I (low range) = 500 mA (3744B), 50 mA
(3724B),
LIM P = 5000 mW (3744B), 1000 mW (3724B), 200
mW (3714B)
LIM V = 10 volts LASER DISPLAY enabled, in I mode
LASER output off LASER display showing actual current (0.00 with
output off)
Constant I, low bandwidth mode selected LASER Tolerance values = 10.0 mA, 1.0 seconds
(3744B),
LASER STEP value = 1 LASER I Set Point = 0 mA
1.00 mA, 1.0 seconds (3724B/3714B) LASER P
PD
Set Point = 0 mW
LASER IPD Set Point = 0 mA
RECALL BIN number = 0