INTRODUCTION AND SPECIFICATIONS
Installing the LDC-3700B
6 LDC-3700 Series
CHAPTER 1
7. Measured at 1 Amp output over a bandwidth of 10 Hz to 10 MHz.
8. When ordered with TSC−599 RTD Sensor Convertor.
9. Nominal temperature coefficients, It and Vt, apply over the rated IC temperature sensor range.
10. Using a 100kW thermistor, controlling an LDM−4412 mount over −30 °C to +25 °C.
11. Using a 10kW thermistor, controlling an LDM−4412 mount over 0 °C to +90 °C.
12. Software limits of display range.
13. Voltage measurement accuracy while driving calibration load. Accuracy is dependent upon load used.
Table 1.2 General LDC−37x4B Specifications
Connectors
Photodiode Monitor and Current Source Connectors: 9−pin, D−connector, for LASER output
External Modulation Connector BNC, instrumentation amplifier input
Temperature Controller: 15−pin D−connector, for TEC output and thermistor
input
GPIB Interface Meets ANSI/IEEE Std 488.1−1987
Meets ANSI/IEEE Std 488.2−1987
General
Size (H x W x D) 5.0" x13.9" x 13.6", 127 mm x 353 mm x 345 mm
Weight
LDC−3714B
LDC−3724B
LDC−3744B
approx 22.5 lbs (10.2 kg)
approx 22.5 lbs (10.2 kg)
approx 25 lbs (11.3 kg)
Power Requirements 100 V ±10%, 120 V ±10%, 220 V ±10%, 230−240 V
±10%, 50−60 Hz
Temperature 0 °C to 40 °C operating; −40 to +70 °C storage
Humidity < 90 % relative humidity, non−condensing
Laser Safety Features: Key switch, interlock and output delay (meets CDRH
US21 1040.10)
LASER Display type: 5−digit, green LED
TEC Display type: 5−digit, green LED
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com