ILX Lightwave LDC-3700B Series - Table 1.1 Temperature Controller Specifications

ILX Lightwave LDC-3700B Series
152 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
INTRODUCTION AND SPECIFICATIONS
Installing the LDC-3700B
7/01 LDC-3700 Series 5
CHAPTER 1
Table 1.1 Temperature Controller Specifications
1
1. All values relate to a one−hour warm−up period.
Temperature Control Output
Temperature Control Range:
2
2. Software limits of range. Actual range possible depends on the physical load, thermistor type, and TE module used.
−100.0 °C to +199.9 °C
Thermistor Set point Resolution and Accuracy:
−20 °C to 20 °C
20 °C to 50 °C
Resolution:
0.1 °C
0.2 °C
Accuracy
3
:
±0.2 °C
±0.2 °C
3. Accuracy figures are quoted for a typical 10 kW thermistor and 100 mA Current setting. Accuracy figures are relative to the calibration
standard. Both resolution and accuracy are dependent upon the user−defined configuration of the instrument.
AD590 & LM335 Set point:
4
−20 °C to 50 °C
4. Accuracy depends on the sensor model selected, the calibration standard, and the user−defined configuration of the instrument.
0.01 °C ±0.1°C
Short Term Stability (1 hr.):
5
5. Over any 1−hour period, half−scale output, controlling an LDM−4412 mount @ 25° C, with 10 kW thermistor, on 100 mA setting.
<0.004 °C
Long Term Stability (24 hrs.):
6
6. Over any 24−hour period, half−scale output, controlling an LDM−4412 mount @ 25° C, with 10 kW thermistor, on 100 mA setting.
<0.01 °C
Output Type: Bipolar, constant current source
Compliance Voltage: >4 V DC
Short Circuit Output Current: 4 Amps
Maximum Output Power: 16 Watts
Current Noise and Ripple:
7
<1 mA rms
Current Limit Range: 0−4 Amps
Current Limit Set Accuracy: ±0.05 Amps
Control Algorithm: Smart Integrator, Hybrid PI
Temperature Sensor
Types:
Thermistor:
IC Temperature Sensor:
RTD Sensor:
8
NTC (2−wire)
AD590/LM335
Pt100/Other 100−ohm RTD
Thermistor Sensing Current: 10 µA/100 µA
Sensor Bias: AD590 = 8V, LM335 = 1 mA, RTD=0.8 mA
8
Useable Thermistor Range: 25−450,000 , typical
Typical Sensor Output
9
AD590 Current Output:
LM335 Voltage Output:
RTD (P
t
100) Resistance:
I(25 °C) = 298.2 µA, It = 1 µA/K
V(25 °C) = 2.73 V, Vt = 10 µV/K
R(25 °C)=109.73
User Calibration: Thermistor: Steinhart−Hart, 3 constants
IC Sensors, RTD = Two−point
TEC Measurement (display)
Temperature:
10 mA Setting:
10
100 µA Setting:
11
Range
12:
−99.9 °C to 199.9 °C
−99.9 °C to 199.9 °C
Resolution:
0.01 °C
0.01 °C
Accuracy:
±0.1 °C
±0.05 °C
Thermistor Resistance
10 µA Setting:
100 µA Setting:
0.01−450.00 k
0.001−45.000 k
0.01 k
0.001 k
±0.05%
±0.05%
TE Current: −4.000 A to 4.000 A 0.001 A ±0.04 A
Voltage (Remote only):
13
−10.000 V to 10.000 V 1 mV ±30 mV
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Related product manuals