CALIBRATION AND TROUBLESHOOTING
Troubleshooting
7/01 LDC-3700 Series 131
CHAPTER 5
Error Messages
In the event of a hardware error condition, error messages are displayed on the TEC or LASER
displays. In most cases, the error message appears for three seconds. In the case of multiple error
messages, the Controller will sequentially show each message for three seconds. In addition to the
hardware errors, GPIB errors may be read via the"ERR?" query.
Table 5.4 lists the numerical error ranges by function. Table 5.5 contains all of the potential error
messages. Not all of these messages may appear on the front panel displays; some refer to GPIB
activities only. ;
In remote operation, the errors can be read by issuing the "ERR?" query. When this is done, all of
the error messages which are resident in the error queue are returned (up to 10 may be stored).
Reading the error queue via GPIB clears the error queue.
Note: Errors codes not listed are reserved for future design use.
Table 5.4 Error Code Classifications
Error Code Range Area of Operation
E−001 to E−099 Internal Program Errors
E−100 to E−199 Parser Errors
E−200 to E−299 Execution Control Errors
E−300 to E−399 GPIB Errors
E−400 to E−499 TEC Control Errors
E−500 to E−599 LASER Control Errors
Table 5.5 Error Messages
Error Code Explanation
E−001 Memory allocation failure.
E−103 <DEFINITE LENGTH ARBITRARY BLOCK PROGRAM DATA> length too long.
E−104 <NON−DECIMAL NUMERIC PROGRAM DATA> type not defined.
E−105 <DECIMAL PROGRAM DATA> exponent not valid.
E−106 <DECIMAL PROGRAM DATA> digit expected.
E−124 <program mnemonic> Lookup, failed because query/command type match failed.
E−126 Too few or too many program data elements.
E−201 <PROGRAM DATA> value out of range.
E−202 <PROGRAM DATA> will not convert to valid type.
E−203 Security violation, command is not available without clearance.
E−205 <PROGRAM DATA> is not a Boolean value or word.
E−207 <PROGRAM DATA> will not convert to an unsigned 16−bit value.
E−210 <PROGRAM DATA> will not convert to a floating point value.
E−211 <PROGRAM DATA> will not convert to a character value.
E−213 <PROGRAM DATA> is incorrect block data length.
E−214 <PROGRAM DATA> length exceeds maximum.
E−301 A <RESPONSE MESSAGE> was ready, but controller failed to read it. (Query error).